Division 8.1.8.4.4 SSV calibration
United Kingdom
A subsonic venturi (SSV) shall be calibrated to determine its calibration coefficient, C d, for the expected range of inlet pressures. An SSV flow meter shall be calibrated as follows:
- (a)
The system shall be connected as shown in Figure 6.5;
- (b)
The blower shall be started downstream of the SSV;
- (c)
Leaks between the calibration flow meter and the SSV shall be less than 0,3 % of the total flow at the highest pressure restriction;
- (d)
While the SSV operates, a constant temperature at the SSV inlet shall be maintained within ± 2 % of the mean absolute inlet temperature, T in;
- (e)
The variable restrictor or variable-speed blower shall be set to a flow rate greater than the greatest flow rate expected during testing. Flow rates may not be extrapolated beyond calibrated values, so it is recommended that it is made certain that a Reynolds number, Re, at the SSV throat at the greatest calibrated flow rate is greater than the maximum Re expected during testing;
- (f)
The SSV shall be operated for at least 3 min to stabilize the system. The SSV shall continue operating and the mean of at least 30 s of sampled data of each of the following quantities shall be recorded:
- (i)
The mean flow rate of the reference flow meter, ;
- (ii)
Optionally, the mean dew point of the calibration air, T dew. See Annex VII for permissible assumptions;
- (iii)
The mean temperature at the venturi inlet, T in;
- (iv)
The mean static absolute pressure at the venturi inlet, p in;
- (v)
Static differential pressure between the static pressure at the venturi inlet and the static pressure at the venturi throat, Δp SSV;
- (i)
- (g)
The restrictor valve shall be incrementally closed or the blower speed decreased to decrease the flow rate;
- (h)
The steps in paragraphs (f) and (g) of this point shall be repeated to record data at a minimum of ten flow rates;
- (i)
A functional form of C d versus Re shall be determined by using the collected data and the equations in Annex VII;
- (j)
The calibration shall be verified by performing a CVS verification (i.e., propane check) as described in point 8.1.8.5 using the new C d versus Re equation;
- (k)
The SSV shall be used only between the minimum and maximum calibrated flow rates;
- (l)
The equations in section 3 of Annex VII (molar based approach) or section 2 of Annex VII (mass based approach) shall be used to determine SSV flow during a test.