Division 8.1.8.4.4 SSV calibration

United Kingdom

Original text. Later amendments are not shown.

A subsonic venturi (SSV) shall be calibrated to determine its calibration coefficient, C d, for the expected range of inlet pressures. An SSV flow meter shall be calibrated as follows:

  1. (a)

    The system shall be connected as shown in Figure 6.5;

  2. (b)

    The blower shall be started downstream of the SSV;

  3. (c)

    Leaks between the calibration flow meter and the SSV shall be less than 0,3 % of the total flow at the highest pressure restriction;

  4. (d)

    While the SSV operates, a constant temperature at the SSV inlet shall be maintained within ± 2 % of the mean absolute inlet temperature, T in;

  5. (e)

    The variable restrictor or variable-speed blower shall be set to a flow rate greater than the greatest flow rate expected during testing. Flow rates may not be extrapolated beyond calibrated values, so it is recommended that it is made certain that a Reynolds number, Re, at the SSV throat at the greatest calibrated flow rate is greater than the maximum Re expected during testing;

  6. (f)

    The SSV shall be operated for at least 3 min to stabilize the system. The SSV shall continue operating and the mean of at least 30 s of sampled data of each of the following quantities shall be recorded:

    1. (i)

      The mean flow rate of the reference flow meter, q – V ref ;

    2. (ii)

      Optionally, the mean dew point of the calibration air, T dew. See Annex VII for permissible assumptions;

    3. (iii)

      The mean temperature at the venturi inlet, T in;

    4. (iv)

      The mean static absolute pressure at the venturi inlet, p in;

    5. (v)

      Static differential pressure between the static pressure at the venturi inlet and the static pressure at the venturi throat, Δp SSV;

  7. (g)

    The restrictor valve shall be incrementally closed or the blower speed decreased to decrease the flow rate;

  8. (h)

    The steps in paragraphs (f) and (g) of this point shall be repeated to record data at a minimum of ten flow rates;

  9. (i)

    A functional form of C d versus Re shall be determined by using the collected data and the equations in Annex VII;

  10. (j)

    The calibration shall be verified by performing a CVS verification (i.e., propane check) as described in point 8.1.8.5 using the new C d versus Re equation;

  11. (k)

    The SSV shall be used only between the minimum and maximum calibrated flow rates;

  12. (l)

    The equations in section 3 of Annex VII (molar based approach) or section 2 of Annex VII (mass based approach) shall be used to determine SSV flow during a test.