OpenLegislation

Paragraph 5.1 F1

The MPE due to an influence factor shall be as specified in Table 10.

Table 10

Load (m) in verification scale intervals (d)MPE
0 < m ≤ 500± 0.5 d
500 < m ≤ 2,000± 1.0 d
2,000 < m ≤ 10,000± 1.5 d

Annotations

Textual amendments

F1

Schs. 1A-1K inserted (E.W.S.) (31.12.2020) by The Product Safety and Metrology etc. (Amendment etc.) (EU Exit) Regulations 2019 (S.I. 2019/696), reg. 1, Sch. 27 para. 49 (with Sch. 27 para. 50(a)) (as amended by S.I. 2020/676, regs. 1(1), 2); 2020 c. 1, Sch. 5 para. 1(1)