Paragraph 7 Multi-register metersI1
England, Wales and Scotland
Induction meters
All induction meters with more than one register shall be tested on one register in accordance with paragraphs 5 and 6 and on each and every other register at a low load using the method A, method B or method C test and at an intermediate or high load using the method A test.
For the same load conditions the maximum permitted difference between the error on one register, expressed as a percentage, and the error on any other register, expressed as a percentage, is 1 per cent.
Static meters
All static meters with more than one register shall be tested on one register in accordance with paragraphs 5 and 6.
For repaired static meters with more than one register—
where the total units are the sum of all the registers, a further test shall be carried out on each and every other register using method A but
where the total units are recorded on one register, only that register is required to be tested in accordance with paragraphs 5 and 6.