Paragraph 2.3 Verification scale interval — multi-interval instrumentsF1

Table 3

1 For i = r the corresponding column of Table 2 applies with e replaced by er.

Verification scale intervalNumber of verification scale intervalsn = Max/e

Minimum value1

n = Maxi /e(i+1)

For i = r the corresponding column of Table 2 applies with e replaced by er.

Maximum value

n = Maxi /ei

Y(I)0.001 g ≤ ei50,000
Y(II)0.001 g ≤ ei≤ 0.05 g5,000100,000
0.1 g ≤ ei5,000100,000
Y(a)0.1 g ≤ ei50010,000
Y(b)5 g ≤ ei501 000

Where:

  • i =  1, 2, … r

  • i =  partial weighing range

  • r =  total number of partial ranges