Division 8.1.8.5.5 Propane check performance
United Kingdom
- (a)
The propane check shall be performed as follows:
- (i)
For batch HC sampling, clean storage media, such as evacuated bags shall be connected;
- (ii)
HC measurement instruments shall be operated according to the instrument manufacturer's instructions;
- (iii)
If correction for dilution air background concentrations of HC is foreseen, background HC in the dilution air shall be measured and recorded;
- (iv)
Any integrating devices shall be zeroed;
- (v)
Sampling shall begin and any flow integrators shall be started;
- (vi)
C3H8 shall be released at the rate selected. If a reference flow rate of C3H8 is used, the integration of this flow rate shall be started;
- (vii)
C3H8 shall be continued to be released until at least enough C3H8 has been released to ensure accurate quantification of the reference C3H8 and the measured C3H8;
- (viii)
The C3H8 cylinder shall be shut off and sampling shall continue until it has been accounted for time delays due to sample transport and analyzer response;
- (ix)
Sampling shall be stopped and any integrators shall be stopped;
- (i)
- (b)
In case the metering with a critical flow orifice is used, the following procedure may be used for the propane check as the alternative method of point 8.1.8.5.5(a);
- (i)
For batch HC sampling, clean storage media, such as evacuated bags shall be connected;
- (ii)
HC measurement instruments shall be operated according to the instrument manufacturer's instructions;
- (iii)
If correction for dilution air background concentrations of HC is foreseen, background HC in the dilution air shall be measured and recorded;
- (iv)
Any integrating devices shall be zeroed;
- (v)
The contents of the C3H8 reference cylinder shall be released at the rate selected;
- (vi)
Sampling shall begin, and any flow integrators started after confirming that HC concentration is to be stable;
- (vii)
The cylinder's contents shall be continued to be released until at least enough C3H8 has been released to ensure accurate quantification of the reference C3H8 and the measured C3H8;
- (viii)
Any integrators shall be stopped;
- (ix)
The C3H8 reference cylinder shall be shut off.
- (i)